The molar absorption coefficient is a sample dependent property and is a measure of how strong an absorber the sample is at a particular wavelength of light.
The concentration is simply the moles L -1 M of the sample dissolved in the solution, and the length is the length of the cuvette used for the absorbance measurement and is typically 1 cm. The Beer-Lambert law states that there is a linear relationship between the concentration and the absorbance of the solution, which enables the concentration of a solution to be calculated by measuring its absorbance.
To demonstrate this linear dependence five solutions of Rhodamine B in water were measured using the DS5 Dual Beam Spectrophotometer Figure 3a and from these absorption spectra, a linear calibration curve of the absorbance versus concentration was created Figure 3b.
Using this calibration curve the concentration of an unknown Rhodamine B solution can be determined by measuring its absorbance which is the main utility of the Beer-Lambert Law. For more information on the theory of absorption spectroscopy, check out the frequently asked questions section on our blog. McNaught and A. Blackwell Scientific Publications Why not browse our range below:. The Beer-Lambert law relates the attenuation of light to the properties of the material through which the light is traveling.
This page takes a brief look at the Beer-Lambert Law and explains the use of the terms absorbance and molar absorptivity relating to UV-visible absorption spectrometry.
For each wavelength of light passing through the spectrometer, the intensity of the light passing through the reference cell is measured. The absorbance of a transition depends on two external assumptions. This formula is the common form of the Beer-Lambert Law , although it can be also written in terms of intensities:.
On most of the diagrams you will come across, the absorbance ranges from 0 to 1, but it can go higher than that. An absorbance of 0 at some wavelength means that no light of that particular wavelength has been absorbed. In a sample with an absorbance of 1 at a specific wavelength, what is the relative amount of light that was absorbed by the sample? You will find that various different symbols are given for some of the terms in the equation - particularly for the concentration and the solution length.
The Greek letter epsilon in these equations is called the molar absorptivity - or sometimes the molar absorption coefficient. The larger the molar absorptivity, the more probable the electronic transition.
In uv spectroscopy, the concentration of the sample solution is measured in mol L -1 and the length of the light path in cm. Thus, given that absorbance is unitless, the units of molar absorptivity are L mol -1 cm However, since the units of molar absorptivity is always the above, it is customarily reported without units.
Guanosine has a maximum absorbance of nm. What is the concentration of guanosine? What is the extinction coefficient? The proportion of the light absorbed will depend on how many molecules it interacts with.
Suppose you have got a strongly colored organic dye. If the path length is known, the slope of the line can then be used to calculate the molar absorptivity. The third step is to measure the absorbance in the sample with an unknown concentration. The absorbance of the sample is used with the equation for the standard curve to calculate the concentration. The way to think about this question is to consider the expression we wrote earlier for the absorbance.
Since stray radiation always leaks in to the detector and presumably is a fixed or constant quantity, we can rewrite the expression for the absorbance including terms for the stray radiation. At low concentration, not much of the radiation is absorbed and P is not that much different than P o.
If the sample is now made a little more concentrated so that a little more of the radiation is absorbed, P is still much greater than P S. As the concentration is raised, P, the radiation reaching the detector, becomes smaller. If the concentration is made high enough, much of the incident radiation is absorbed by the sample and P becomes much smaller.
At its limit, the denominator approaches P S , a constant. The ideal plot is the straight line. Spectroscopic instruments typically have a device known as a monochromator. There are two key features of a monochromator. The first is a device to disperse the radiation into distinct wavelengths.
You are likely familiar with the dispersion of radiation that occurs when radiation of different wavelengths is passed through a prism. The term effective bandwidth defines the packet of wavelengths and it depends on the slit width and the ability of the dispersing element to divide the wavelengths. The important thing to consider is the effect that this has on the power of radiation making it through to the sample P o. Reducing the slit width will lead to a reduction in P o and hence P. An electronic measuring device called a detector is used to monitor the magnitude of P o and P.
All electronic devices have a background noise associated with them rather analogous to the static noise you may hear on a speaker and to the discussion of stray radiation from earlier that represents a form of noise. P o and P represent measurements of signal over the background noise. As P o and P become smaller, the background noise becomes a more significant contribution to the overall measurement. Ultimately the background noise restricts the signal that can be measured and detection limit of the spectrophotometer.
Therefore, it is desirable to have a large value of P o. Since reducing the slit width reduces the value of P o , it also reduces the detection limit of the device. Selecting the appropriate slit width for a spectrophotometer is therefore a balance or tradeoff of the desire for high source power and the desire for high monochromaticity of the radiation.
It is not possible to get purely monochromatic radiation using a dispersing element with a slit. Usually the sample has a slightly different molar absorptivity for each wavelength of radiation shining on it. The net effect is that the total absorbance added over all the different wavelengths is no longer linear with concentration. Instead a negative deviation occurs at higher concentrations due to the polychromicity of the radiation.
Furthermore, the deviation is more pronounced the greater the difference in the molar absorbtivity. As the molar absorptivities become further apart, a greater negative deviation is observed.
Therefore, it is preferable to perform the absorbance measurement in a region of the spectrum that is relatively broad and flat. The peak at approximately nm is quite sharp whereas the one at nm is rather broad. Given such a choice, the broader peak will have less deviation from the polychromaticity of the radiation and is less prone to errors caused by slight misadjustments of the monochromator. It is important to consider the error that occurs at the two extremes high concentration and low concentration.
A relatively small change in the transmittance can lead to a rather large change in the absorbance at high concentrations. At very low sample concentrations, we observe that P o and P are quite similar in magnitude.
If we lower the concentration a bit more, P becomes even more similar to P o. The important realization is that, at low concentrations, we are measuring a small difference between two large numbers. For example, suppose we wanted to measure the weight of a captain of an oil tanker.